-
NT & RP Journal
PROPOSED VALUES FOR DIAGNOSTIC REFERENCE LEVELS IN INTRAORAL DENTAL RADIOGRAPHY IN SERBIA
......: info......: history......: editorial......: archive......: for authors......: subscription
   
Image NTRP

Vol. XXXIX, No. 1, Pp. 1-87
March 2024
UDC 621.039+614.876:504.06
ISSN 1451-3994

Back to Contents

 


Pages: 66-73

Authors:
Zoran M. Mirkov, Nebojša T. Milošević, and Dario Faj

Abstract

Values for diagnostic reference levels are proposed for intraoral dental radiography in the Republic of Serbia. Proposed numbers for diagnostic reference levels were based on measurements on 119 intraoral units in clinical settings for an adult posterior bitewing X-ray. Values of 3.17 mGy, 3.11 mGy, and 1.58 mGy incident air kerma were found for image receptors film E class, charge-coupled device, and photostimulable phosphor plate, respectively. Similarly, 89.6 mGycm2, 88 mGycm2, and 44.6 mGycm2 air kerma-area products were found for the same detectors. All values are greater than those published in several other similar studies. This is especially true for charge-coupled device image receptors since this technology enables lower patient doses as reported before. This calls for urgent action to optimize intraoral dental imaging, so we present the radiographic techniques used and equipment description to help guide optimization actions.

Key words: dental radiography, radiation protection, diagnostic reference level

FULL PAPER IN PDF FORMAT (1,09 MB)

Vinča Institute of Nuclear Sciences :: Designed by milas :: July 2007
Operated by acapanic :: Last updated on July, 2024