Vol.
XXXIV, No. 4, Pp. 313-398
December 2019
UDC 621.039+614.876:504.06
ISSN 1451-3994
Back to Contents
|
Pages: 353-360
Authors: Mona M. GOUDA
Abstract
In this article, the full energy peak efficiency of NaI detector using non-axial cylindrical sources is calculated by using a new efficient theoretical approach. This approach depends on using the efficiency transfer method and analytical calculations of the average path length of a gamma photon inside the source to the detector system. Measured efficiencies made by using 152Eu aqueous radioactive cylindrical sources with volumes 25 ml and 400 ml. Comparing calculated efficiencies to the measured one showed good agreement enabling the validation of this approach.
Key words: full energy peak efficiency, cylindrical source, average path length,
effective solid angle ratio, efficiency transfer method
FULL PAPER IN PDF FORMAT (985 KB)
|