CALIBRATION OF NaI (Tl) CYLINDRICAL DETECTOR USING AXIALLY SHIFTED RADIOACTIVE CYLINDRICAL SOURCES
......: info......: history......: editorial......: archive......: for authors......: subscription
   
Image NTRP

Vol. XXXIV, No. 4, Pp. 313-398
December 2019
UDC 621.039+614.876:504.06
ISSN 1451-3994

Back to Contents

 


Pages: 353-360

Authors: Mona M. GOUDA

Abstract

In this article, the full energy peak efficiency of NaI detector using non-axial cylindrical sources is calculated by using a new efficient theoretical approach. This approach depends on using the efficiency transfer method and analytical calculations of the average path length of a gamma photon inside the source to the detector system. Measured efficiencies made by using 152Eu aqueous radioactive cylindrical sources with volumes 25 ml and 400 ml. Comparing calculated efficiencies to the measured one showed good agreement enabling the validation of this approach.

Key words: full energy peak efficiency, cylindrical source, average path length,
effective solid angle ratio, efficiency transfer method

FULL PAPER IN PDF FORMAT (985 KB)