NT & RP Journal
A SURVEY OF SHORT-TERM AND LONG-TERM STABILITY OF TUBE PARAMETERS IN A MAMMOGRAPHY UNIT
......: info......: history......: editorial......: archive......: for authors......: subscription
image

Vol. XXIX, No. 4, Pp. 253-331
December 2014
UDC 621.039+614.876:504.06
ISSN 1451-3994

Back to Contents

 


Pages: 321-325

Authors:
Jovica Ž. Praskalo, Jasna Dj. Davidović, Biljana V. Kočnić, Monika M. Živković,
and Svetlana M. Pejović


Abstract

In order to set up a successful mammography screening program in the Republic of Srpska, a Siemens Mammomat 1000 X-ray machine was selected for analysis as the said mammography system is widely used in clinical practice. The variations in tube parameters (specific air kerma, high-voltage accuracy and reproducibility, linearity between exposure and dose exposure time) were monitored over a five-year period, from 2008 to 2012. In addition, due to observed daily fluctuations for chosen parameters, a series of measurements were performed three times a day within a single-month period (mainly October 2012). The goal of such an experimental set up is to assess short-term and long-term stability of tube parameters in the given mammography unit and to make a comparison between them. The present paper shows how an early detection of significant parameter fluctuations can help eliminate irregularities and optimize the performance of mammography systems.

Key words: mammography, air kerma, quality control

FULL PAPER IN PDF FORMAT (586 KB)

Vinča Institute of Nuclear Sciences :: Designed by milas :: July 2007
Operated by acapanic :: Last updated on February, 2015