Abstract


In this paper, basic requirements for designing of a web Laboratory were given briefly. Proposed requirements are flexible and independent of hardware used, software, scalability, and area of application, i. e. of the type of experiment. System enables access to the experiment using web browser, experimental parameters setup, start of the experiment in the real time and review of the obtained results in numerical and graphical form. Besides, original software for remote access and control of Atomtex AT1125A instrument is presented in the paper. The software enables that data measured by Atomtex AT1125A to be given in the format which is suitable for further processing and also enables that activities such as reading, collecting, and processing data to be carried out automatically or manually, according to the given algorithm (time of reading, alarm for the given dose level, statistical error). Remote access to Atomtex 1125A instrument, realized through web application, enables radiation safety of personnel during long, risky, and difficult conditions of measurements.

Key words: Atomtex AT1125A  instrument, remote access, radiation safety