NT & RP Journal

BETA-EFFICIENCY OF A TYPICAL GAS-FLOW IONIZATION CHAMBER USING GEANT4 MONTE CARLO SIMULATIONS

......: info......: history......: editorial......: archive......: for authors......: subscription

Vol. XXVI, No. 3, Pp. 181-274
December 2011
UDC 621.039+614.876:504.06 ISSN 1451-3994

Back to Contents

 



Pages:
193-200

Authors:
Abid Hussain, Sikander M. Mirza, Nasir M. Mirza, Muhammad T. Siddique

Abstract

GEANT4 based Monte Carlo simulations have been carried out for the determination of effi -
ciency and conversion factors of a gas-flow ionization chamber for beta particles emitted by
86 different radioisotopes covering the average-β energy range of 5.69 keV-2.061 MeV. Good
agreements were found between the GEANT4 predicted values and corresponding experi -
mental data, as well as with EGS4 based calculations. For the reported set of β-emitters, the
values of the conversion factor have been established in the range of 0.5·1013-2.5·1013
Bqcm-3/A. The computed xenon-to-air conversion factor ratios have attained the minimum
value of 0.2 in the range of 0.1-1 MeV. As the radius and/or volume of the ion chamber in -
creases, conversion factors approach a flat energy response. These simulations show a small,
but significant dependence of ionization efficiency on the type of wall material.

 

 

Key words: ionization chamber, gas-flow, efficiency, GEANT4 simulations

FULL PAPER IN PDF FORMAT ( 917 KB)

Vinča Institute of Nuclear Sciences :: Designed by milas :: July 2007
Last updated on January, 2012